Improving Lives Through Science
›Resources› Energy-saving Features of Esco Frontier® Acela Ducted Fume Hood
Esco Frontier® Acela are high-performance fume hoods that can operate at 60 fpm instead of traditional 100 fpm that saves about $2000/year in electric bills from lower exhaust and make-up air requirement (heating and cooling).The small additional price for a high-performance fume hood will pay off in the first year of usage when compared to a conventional fume hood.
Energy Efficiency
Fume hoods are essential laboratory safety devices that are highly energy-intensive, each one consuming more energy than three homes in an average U.S.climate.Depending on climate and system design, estimated energy costs for fume hoods range up to US$9,000 annually1, based on face velocities of 0.5 m/s (100 fpm) at full sash open position for a 1.8 m (72.0") hood.
Variable Air Volume (VAV) is one of the various approaches presently employed to reduce hood energy consumption.The table below compares conventional hoods, VAV hoods, and the Esco Frontier® Acela high-performance low flow hood.
Table 1.Comparison of Conventional hoods, VAV hoods, and Esco Frontier® Acela ducted fume hoods
1Energy use and savings potential for laboratory fume hoods.Evan Mills, Dale Sartor!Energy, 2003
Compared with conventional hoods, Esco Frontier® Acela operates safely at 0.3 m/s (60 fpm) at 457 mm (18.0") or full open sash position while maintaining excellent ASHRAE and EN containment.Exhaust volume reductions of up to 60% may be achieved without compromising safety.This translates into an annual operating cost savings of up to US$5,600.Unlike VAV systems, the Esco Frontier® Acela is easy and inexpensive to install, commission, and maintain.
Table 2.Exhaust Volume Reductions (Esco Frontier® Acela fume hood vs.Conventional fume hood)
High containment is achieved by incorporating aerospace technology using airfoils and Computational Fluid Dynamics:
1.Airfoils on all 4 front opening perimeters, to create high velocity airflow to prevent chemical vapor leak on this critical area, where the airflow is weak on conventional hood.
a.Sash Handle Airfoil
b.Side Airfoil
c.Armrest Airfoil
2.Variable perforated back wall to even out suction and make the airflow unidirectional, with added benefit of reducing chance for cross-contamination between one side to another side of the hood.
a.Back Baffle
3.Slots in 4 different heights to quickly remove light, medium, heavy weight chemical vapor, and top bypass to reduce vortex on top.
a.Four different height chemical vapor removal, depends on vapor weight:
- light vapor (top)
- medium-light (second from top)
- medium (middle and back perforation)
- heavy (bottom)
4.Evenly distributed exhaust suction, using trapezoidal exhaust collar and variable back perforation to ensure that there is no vapor leak from left and right walls, which are the furthest points from the blower.
a.Trapezoidal exhaust collar and larger back perforations on the left / right sides to distribute the exhaust suction evenly across work zone width:
Esco Frontier® Acela (EFA)a high-performance, low-flow, ducted fume hood designed with precisely tuned aerodynamics, outstanding user safety, design flexibility, energy efficiency, and aesthetics.A conventional laboratory fume hood that lets you save while keeping you safe.
Contact us:
Esco Technologies, Inc.
Scientific Division
903 Sheehy Drive, Suite F, Horsham, PA 19044, USA
Toll-Free USA and Canada: 1-877-479-3726
215-441-9661
484-698-7757
eti.admin@escolifesciences.com